Signet 2610 Process Optical Dissolved Oxygen Sensor3-2610.091 Rev. A 04/12 English*3-2610.091*EnglishMODBUS ManualCable length must not exceed 1220
22610 DO Sensor MODBUS ManualRegisterSize(registers)Mode & Access Level (R/W)DataTypeDescriptionDissolved Oxygen Concentration40038 2 R1 fl oat Mea
32610 DO Sensor MODBUS ManualLive Salinity ValueThe live salinity value is used to correct the oxygen concentration value for salinity. Values must be
42610 DO Sensor MODBUS ManualCalibrate the sensor using the following procedure:1. Optional: Read the Sensor Data Cache Timeout register 49463 and st
Calibration CalculationsCalibrated oxygen reading:O2RC = C0 + C1 × O2RUWhere:C1 = (O2100%Sat)/(O2RUS × O2RUZ)C0 = –C1 × O2RUZWhere:O2100%Sat is the th