Signet 8900 Multi-Parameter ControllerMulti-ParameterControllerC1 2.50 µS/cmF2 58.43 GPMRelay 1Relay 2ENTER3-8900.090-13-8900.090-1 Rev. K 01/13
10Signet 8900 Multi-Parameter5. Mounting the Base UnitThe 8900 may be mounted in a panel, on a wall, or on virtually any surface including shelves, r
11 Signet 8900 Multi-Parameter5.2 Mounting AccessoriesMfr. Part No. Description3-8050.395 Splashproof rear cover • Use in conjunction with liq
12Signet 8900 Multi-Parameter6. Wiring All wiring connections to the 8900 are made via the removable terminals of the plug-in modules. This section c
13 Signet 8900 Multi-Parameter6.2 Digital (S3L) Input wiring• The maximum cable length of the digital (S3L) bus varies depending on the types of sens
14Signet 8900 Multi-ParameterStep 1: Calculate the Total current requirements for digital (S3L) BranchesThis information will determine the total cur
15 Signet 8900 Multi-Parameter3-8900.621CI/O Module 3-8900.401-X1234567891011121314+5VDC (Black)Freq. Input (Red)GND (Shield)+5VDC (Black)Freq. Input
16Signet 8900 Multi-Parameter3-8900.621CI/O Module 3-8900.401-X1234567891011121314+5VDC (Black)Freq. Input (Red)GND (Shield)+5VDC (Black)Freq. Input 2
17 Signet 8900 Multi-Parameter6.3 Power Module (3-8900.402-X)• The 8900 can operate from a Universal AC power input (3-8900.402-1) or from a 24 VDC
18Signet 8900 Multi-Parameter l r er ale r er M6.5 Relay Module
19 Signet 8900 Multi-ParameterF1 123.45 GPMP3 4.32 pHF1 123.45 GPMF2 54.321 GPMRatio F1:F22.27P3 4.32 pHL4 19.49 cmRst1 1234567.8Gallons >Perm1 123
2Signet 8900 Multi-ParameterGeneral Number of input channels: 2 or 4 or 6Input signal types:Digital (S3L): Serial ASCII, TTL level 9600 bps Freque
20Signet 8900 Multi-Parameter7.3 Menu Directory• Select this item to exit View and go to the Menu Directory.• The Menu Directory may also be reach
21 Signet 8900 Multi-ParameterMenu Directory:System SetupReturning toViewMode...Channel 1 Type: None >Channel 1 Type:NoneChannel 1 Type:
22Signet 8900 Multi-ParameterpH ChannelThe default pH display shows the primary pH measurement on line 1 and the secondary temperature measurement on
23 Signet 8900 Multi-Parameter8.5 Other Automatic DisplaysThe 8900 will confi gure the following displays when appropriate:• For each channel assign
24Signet 8900 Multi-Parameter9.1 Channel Settings: If Channel Type = Flow: Set K-Factor (pulses per unit volume) from Flow Sensor manual:Minimum: 0.0
25 Signet 8900 Multi-Parameter9.2 Channel Settings: If Channel Type = pH T1 Units: C >Select the unit of measure for the temperat
26Signet 8900 Multi-ParameterSelect the unit of measure for this conductivity channel:uS/cm mS/cm k • cm M • cm PPM or PPB (of TDS)If un
27 Signet 8900 Multi-ParameterSelect the unit of measure for this pressure channel : psi bar kPa P1 Units: psi >Select decimal locati
28Signet 8900 Multi-Parameter9.7 Channel Settings: If Channel Type = LevelSelect the unit of measure for this Level channel: cm m in ftYes: Le
29 Signet 8900 Multi-ParameterSelect the decimal location for the Volume display. Must be consistent with the range: XXXXX. XXXX.X XXX.XX
3 Signet 8900 Multi-ParameterAnalog Outputs (via I/O Modules and Output Modules)4 to 20 mA Output:Endpoints are adjustable and reversible:Minimum:
30Signet 8900 Multi-ParameterHolds status of input, output and relay functions (fl ow totalizers will continue counting, however). Useful during calib
31 Signet 8900 Multi-Parameter11. Relay Mode DescriptionsThe 8900 offers a wide selection of relay operating modes that enable the user to create com
32Signet 8900 Multi-Parameter105ela utut ate0 t 00 PuleminenintPulse rate0 pulses100 pulsesStarting intProportional Pulsing• Proport
33 Signet 8900 Multi-Parameter11.1 Relay Setup Menu Relay 1 Source: Flow F1 >Select measurement type or derived function to be assigned to
34Signet 8900 Multi-Parameter11.2 Multiple Relay Sources and Advanced Relay ModeWhen a Relay Source is set to "MULTIPLE" and the Relay Mode
35 Signet 8900 Multi-ParameterSelect measurement type or derived function to be assigned to this output.If I/O Module with Voltage Output is installed
36Signet 8900 Multi-Parameter13. CalibrationVolumetric calibrationSelect VOLUMETRIC CALIBRATION if the fl ow rate can be determined by fi lling a vessel
37 Signet 8900 Multi-ParameterSENSOR: Select to perform calibration at the sensor via 2750 EasyCal.INSTRUMENT: Select to perform calibration at th
38Signet 8900 Multi-Parameter13.4 Calibration: If Channel type = ConductivitySet Conductivity C1: >If C~ Cal at Instrument is selected,
39 Signet 8900 Multi-ParameterAdjust LCD contrast: 4 settings.If two or more measurements of the same type are present, select a functional relations
4Signet 8900 Multi-Parameter2. DescriptionThe 8900 Multi-Parameter Controller takes the concept of modularity to the extreme. The controller comes wit
40Signet 8900 Multi-Parameter• % Recovery is a derived measurement based on fl ow rate only. In a reverse osmosis system, % Recovery is fl ow rate in
41 Signet 8900 Multi-ParameterZero reference point (Z):Defi nition: The point in the vessel where you want the 8900 to display zero (0 ft, 0 gal. etc)
42Signet 8900 Multi-ParameterCustom point #103000 gal.Custom point #10 gal.Custom point #245 gal.Custom point #3118 gal.Custom point #4Custom point #5
43 Signet 8900 Multi-ParameterTDS FactorSome industries need to display a conductivity value as Total Dissolved Solids (TDS), measured in units of par
44Signet 8900 Multi-ParameterAppendix D: USP LimitsUSP (United States Pharmacopoeia) has defi ned a set of conductivity values (limits) to be used for
45 Signet 8900 Multi-ParameterAppendix E: Cloning using Digital (S3L) (Serial Cloning)Cloning enables copying the programming from one unit into
46Signet 8900 Multi-ParameterStep 4: Confi gure the Transmitting Controller1. Now go to the Transmitting Controller’s Option menu and scroll until yo
47 Signet 8900 Multi-ParameterAppendix F: BTU Quick Start GuideThis guide outlines how to confi gure the 8900 Multi-Parameter to calculate Power and E
48Signet 8900 Multi-ParameterAppendix G: 8900 display messages8900Multi-ParameterSavingReturning toMenu DirectoryReturning toViewMode...Enter Passwor
49 Signet 8900 Multi-ParameterChannel 1 ID:Not Assigned?Input ModuleNot FoundInput/Output 1&2Module DefectiveComm/Out 3&4Module DefectiveCal P
5 Signet 8900 Multi-Parameter2.2 System OverviewThe most basic 8900 system consists of a Base Unit, an I/O Module and a Power Module. Outputs and rel
50Signet 8900 Multi-ParameterNotes
51 Signet 8900 Multi-ParameterNotes
Ordering InformationMfr. Part No. Code DescriptionBase Unit, required; choose one3-8900 159 000 868 Base unit with back-lit LCDI/O Module,
6Signet 8900 Multi-Parameter3. Installation and Basic FunctionalityThis section lists all items included with individual 8900 system components, descr
7 Signet 8900 Multi-Parameter4.1 I/O Module, required; install one (1) Mfr. Part No. Description3-8900.401-1 Two inputs, no outputs3-8900.401-
8Signet 8900 Multi-Parameter4.3 Output Module, optional; install zero (0) or one (1)Mfr. Part No. Description3-8900.405-1 Two passive 4 to 20
9 Signet 8900 Multi-Parameter8059-4NC COM NORELAY ANC COM NORELAY BNC COM NORELAY CNC COM NORELAY DDC OUTPUT+ -DC INPUT- + + S3L -+